Sync: Label(s) XXX Exist on both sides but are of different types. It is not possible to synchronize

Modified on Tue, 8 May, 2018 at 4:43 PM

1) Review an exam that has been uploaded to the server (Uploaded State)

2) Select Settings Tab, Montage Tab and then "Other Signals" sub-tab (Ensure at the bottom of the page displays "Server Settings")

3) Select Label XXX and write down the "Type"

4) Review an exam that is local to the Acquire (Recording or Recorded State)

5) Select Settings Tab, Montage Tab and then "Other Signals" sub-tab (Ensure at the bottom of the page displays "Local Settings")

6) Select XXX and write down the "Type"

7) Note the "Type" in Steps 6 must be set to Step 3.

8) Close all iEEG Interfaces

9) Open Acquire and Sync the target Perspective


Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article